You are here: Home > News > 维修资讯

东芝CT机spellman电源维修CT42 CT70北京

2019-8-20 13:58:42

东芝CT机spellman电源维修CT42 CT70北京

SPELLMAN高压电源广泛应用于东芝CT机XVISION/EX、AUKLET系列、ASTEION系列、以及多排系列。故障代码及诊断方法大致相同,部分故障代码排查参考如下

Error contents (note 1)

Cause of problem

Object part

24kW

48kW/60kW

E13

Arc-

Cathode side Arc of the X-ray tube or the high-voltage section

TUBE, INV/HV

TUBE, INV/HVC

E25

Arc+

Anode side Arc of the X-ray tube or the high-voltage section

TUBE, INV/HVA

E12

Over kV-

Cathode side overvoltage. Abnormal main circuit control.

INV/HV, SCB

INV/HVC, SCB

E27

Over kV+

Anode side overvoltage. Abnormal main circuit control.

INV/HVA, SCB

E62

Over mA-

Cathode side overcurrent of the X-ray tube or the high-voltage section.

TUBE, INV/HV, SCB

TUBE, INV/HVC,SCB

E63

Over mA+

Anode side overcurrent of the X-ray tube or the high-voltage section.

TUBE, INV/HVA,SCB

E66

MAC error

Filament current abnormal. Trouble of the filament heatig circuit.

INV/HV, SCB

INV/HVC, SCB

E0

Regulation error (HV abnormal)

There is a difference in binding output and real output

INV/HV, AC/C

INV/HVC, INV/HVA, AC/C

spellman电源维修CT机高压电源维修CT70‌‌PN50X4218北京

医疗设备高压电源维修高压发生器维修北京天浦正达高压电源专业维修

Spellman高压电源XRM/PCM50/70维修射频电源北京

Spellman高压电源XRM50P50X3768维修射频电源北京

spellman高压电源维修高压发生器控制单元维修北京

医疗设备高压电源维修高压发生器维修北京天浦正达高压电源专业维修